Bridge power supply
|
DC12 V±0.3 V within 140 mA (Changeable to 3 VDC or 6 VDC)
with remote sensing function
|
Input sensitivity
|
0.3 μV/d or more (d = minimum scale)
|
Input range
|
-3.1 mV/V ~ 3.1 mV/V
|
Zero adjustment range
|
-2.5 mV/V ~ 2.5 mV/V
|
Non-linearity
|
0.01 %F.S.
|
Temperature coefficient
|
Zero point ±0.1 μV/℃, Sensitivity ±0.000 8 %F.S./℃
(When the calibration is set at 0.3 μV/d or more for input sensitivity.)
|
Noise in input signal
|
±0.2 μVP-P or less (At the default setting for analog filter, digital filter, stabilization filter and A/D sampling.)
|
Analog filter
|
Approx. 4 Hz (Changeable to 2 Hz, 6 Hz, 8 Hz, 10 Hz or 100 Hz)
|
A/D sampling speed
|
200 times/s (Changeable to 1 000 times/s, 100 times/s, 50 times/s, 20 times/s or 10 times/s.)
|
A/D internal resolution
|
24 bit
|
Display unit
|
Wide viewing angle 6.2 inches TFT color liquid crystal display
No. of dots : 640 x 240 dots, Dot pitch : 0.077 mm x 0.231 mm
|
Back light
|
Brightness : Approx. 240 cd/m2 (typ.)
Durability : Approx.40 000 hour (typ.) at 25 ℃ until the surface brightness reaches 50 % of initial.
|
Display range
|
- 999 999 ~ 999 999
|
Status display
|
GROSS / NET, TARE, PRE.TARE, STABLE, HOLD, Z-ERR, RUN, CZ,
F.FLOW/S1, M.FLOW/S2, D.FLOW/S3, OVER/S4, OK, UNDER, FULL, Z-BAND, FINISH, FLOW / DISCH., S-COMP / SEQ. / 4-COMP / MEAS. / OUT / MIX
|
Dsiplay of setting value
|
Final, Preliminary-1, Preliminary-2, Over, Under, Free fall
|
Display of brand
|
Code No.: 0 ~ 99, Brand name: Up to 15 characters
|
Display times
|
4 times/s (Changeable to 20 times/s)
|
Display of decimal point
|
Changeable to no display, 101, 102, 103, 104.
|
Accumulation display
|
Accumulation value: 0 ~ 9 999 999 999,
Accumulation times: 0 ~ 999 999
|
Setting method of operation
|
Setting by the operation in Analog resistive touch panel, Also, setting can be made by the communication through RS-485 interface.
|
Various function
|
|
Zero tracking
|
Stabilize the zero shift in a constant condition.
|
Digital filter
|
Stabilize the data by operating process in CPU.
|
Stabilization filter
|
Stabilize by strengthened the digital filter only when the change width of the load is constant.
|
Calibration LOCK switch
|
Lock the calibration function by this switch.
|
Digital linearize
|
Compensate the non-linearity by 3 points at the maximum.
|
Accumulating function
|
Accumulate the loading data, and memorize the times and total value.
|
Preset tare
|
Execute the tare weight cancellation by the setting value input in digital.
|
Operating temp./humidity range
|
Temperature 0 ℃ ~ 40 ℃, Humidity 85 % or less
(Non condensing)
|
Power supply voltage
|
100 VAC ~ AC240 VAC
(Permitted variable range: 85 VAC ~ 264 VAC) 50/60 Hz
|
Power consumption
|
Approx. 20 VA (without any options, at 100 VAC)
Approx. 30 VA or less (with options, at 100 VAC)
|
Outline dimensions(WxHxD)
|
192 mm x 96 mm x 132 mm (excludes protruding parts)
|
Weight
|
Approx. 1.6 kg (without options)
|
Protection code
|
IP65 (Only for the fron panel when the attached panel mounting gascket is installed.)
|